Common Instrument Problems

If you are having trouble with Kingfisher instruments, Please check the following:

An instrument won't respond, or is generally malfunctioning:

An optical power level seems to be wrong or unstable:

An optical power reading seems low at high power levels:

If measuring transmission equipment power levels, it is possible for signal modulation to cause a false reading. This problem is worse at high power levels, and diminishes rapidly as the power level drops. To observe (and cure) this, attenuate the signal into the meter (about 10 dB is good for a trial), and see if the reading comes to the expected value.

A laser source doesn't seem very stable :

An attenuation value seems wrong:

Refer to the meter manual to find out how to take a reference value correctly. Refer to our Application Notes if you are not sure how to use reference values during attenuation measurements.

The instrument is suspected of malfunctioning

Instrument manuals can be downloaded from this web site. In the back of most instrument user manuals, there is a detailed Performance Verification Test which can usually be performed with quite simple equipment. If this test fails, please contact a service center. Any test results you have available will help identify the problem.

I need help running KITS™ Software

Please visit the KITS™ FAQs for common problems, or take a look at the help file for your version of KITS™ software by clicking the help menu item.